10A NCAC 15 .0204         FACILITY RESPONSIBILITIES

(a)  All forms required by this Rule shall be completed in accordance with Rule .0203 of this Section and any accompanying instructions.

(b)  Shielding design requirements:

(1)           Prior to construction for all new installations of X-ray machines for human, non-human, or veterinary use and prior to structural modification of existing installations, an applicant shall have the floor plans, shielding specifications, and equipment arrangement reviewed by a registered service provider.

(2)           The applicant shall submit a shielding design and the agency Shielding Design Review Form to the agency for review. Shielding design drawings shall be scaled. The Shielding Design Review Form shall include the following information:

(A)          facility and service provider name, registration number, e-mail and physical address, and phone number;

(B)          machine location, manufacturer, status, kVp, mA, mA min per week, facility type; and

(C)          proposed date of installation or structural modification.

(3)           An X-ray machine shall not be installed until the applicant has received acknowledgment of the shielding design from the agency.

(4)           An X-ray machine shall not be replaced until the existing shielding design, acknowledged previously by the agency, is reviewed by a registered service provider in accordance with Rule .0205. The registrant shall have a service provider review the acknowledged shielding design for the proposed X-Ray machine replacement to assess if the existing shielding meets the requirements of this Chapter. The documentation provided to the registrant from the service provider shall be submitted to the agency and maintained for agency review during inspection.

(5)           The acknowledgment of such plans shall not preclude the requirement for additional modifications should a subsequent analysis of operating conditions indicate the possibility of a radiation dose that exceeds the limits incorporated in Rule .1601 of this Chapter.

(6)           Shielding designs are not required to be submitted for the following radiation machines:

(A)          dental handheld;

(B)          dual X-ray absorptiometry (DEXA);

(C)          mammography; or

(D)          mobile or portable radiographic and fluoroscopic machines used in more than two locations.

(c)  Facility registration:

(1)           Mobile X-ray machines located and used in this State, and permanently affixed within a vehicle or trailer, shall meet the following requirements prior to use:

(A)          have a shielding design acknowledged by the agency in accordance with Paragraph (b) of this Rule;

(B)          have an X-ray Machine Application form submitted in accordance with Rule .0203(d) of this Section. X-ray machines leased or on loan from a registered service provider shall be registered with the agency if maintained for more than 30 days;

(C)          have a copy of the operating and safety procedures describing how to protect patients, operators, and the public from radiation submitted to the agency; and

(D)          an individual responsible for business operations shall ensure that X-ray machines are operated in accordance with Section .0600 of this Chapter.

(2)           Mobile X-ray machines located out-of-state and brought into this State for use, and permanently affixed within a vehicle or trailer, shall meet the following requirements:

(A)          have the required documentation in Parts (c)(1)(A) through (c)(1)(C) of this Rule submitted as a complete document for agency review;

(B)          have a written notice submitted, in accordance with Rule .0208 of this Section, and maintain it for agency review during inspection; and

(C)          an individual responsible for business operations shall ensure that X-ray machines are operated in accordance with Section .0600 of this Chapter.

(3)           Mobile radiation generating devices located out-of-state and brought into this State for use shall meet the following requirements prior to use:

(A)          have a Radiation Generating Devices Application form submitted in accordance with Rule .0203(d) of this Section. Radiation generating devices leased or on loan from a registered service provider shall be registered with the agency if maintained for use for more than 30 days;

(B)          have a copy of the operating and safety procedures describing how to protect operators and the public from radiation submitted to the agency;

(C)          submit a written notice, in accordance with Rule .0208 of this Section, and maintain it for agency review during inspection; and

(D)          an individual responsible for business operations shall ensure operators are qualified in accordance with Rule .0800 of this Chapter to use the radiation generating device indicated on the application.

(4)           X-ray machines for diagnostic imaging, non-diagnostic imaging, clinical studies, veterinary use, or for use by service providers for demonstration purposes shall meet the following additional requirements:

(A)          have a shielding design acknowledged by the agency in accordance with Paragraph (b) of this Rule;

(B)          submit an X-ray Machine Application form in accordance with Rule .0203(d) of this Section within 30 days of use; and

(C)          an individual responsible for business operations shall ensure that X-ray machines are operated in accordance with Section .0600 of this Chapter.

(5)           Radiation generating devices covered in Section .0800 of this Chapter shall meet the following additional requirements prior to use:

(A)          submit a Radiation Generation Device Application form in accordance with Rule .0203(d) of this Section; and

(B)          an individual responsible for business operations shall ensure operators are qualified in accordance with Rule .0800 of this Chapter to use the radiation generating device indicated on the application.

(6)           Industrial radiography X-ray machines covered in Section .0500 of this Chapter shall meet the following additional requirements prior to use:

(A)          submit a Radiation Generating Device Application form in accordance with Rule .0203(d) of this Section; and

(B)          an individual responsible for business operations shall ensure operators are qualified in accordance with Section .0500 of this Chapter to use the machines indicated on the application.

(d)  Facilities registered pursuant to Paragraph (c) of this Rule shall notify the agency, using the Disposal of an X-Ray Machine or Radiation Generating Device Form, prior to the disposition or the transfer of a registered X-ray machine or radiation generating device to another person required to be registered pursuant to Paragraph (c) of this Rule.

(e)  Facilities registered pursuant to Paragraph (c) of this Rule shall prohibit any person from furnishing services described in Rule .0205(d) of this Section, at his or her facility, until such person provides evidence they are currently registered with the agency as a provider of such services in accordance with Rule .0205 of this Section.

(f)  No facility registered pursuant to the provisions of Paragraph (c) of this Rule shall perform any services listed in Rule .0205(d) of this Section in his or her facility unless such person meets the requirements in Rules .0205 and .0206 of this Section and has received written authorization from the agency to perform such services.

 

History Note:        Authority G.S. 104E‑7; 104E-9(a)(3); 104E-12;